An integrated fluorimeter is required to detect and monitor the levels of fluorescence during the
PCR process for
real-time PCR and there is a range of options available for both the excitation light source and fluorescent emission detection. The light sources that cause fluorophore excitation can be classed as narrow- or broad-spectrum. If a broad-spectrum light source is employed (e.g. Mx4000, Mx3000P, Mx3005P, MyiQ, iQ5, ABI 7300, ABI 7500, LightCycler 480) then filters can be used to provide light tuned to the excitation spectrum of a specific individual fluorophore. Such a system provides the user with a wider choice of available fluorophores, although it is best to select those with good separation of their emission spectra.
A disadvantage of this optical system is that the light intensity passing through the filters can be limited and this could in theory limit the sensitivity of detection. There are currently two narrow-spectrum light sources used in real-time platforms, these can be light emitting diodes (LEDs) (e.g. LightCyclers 1.5 and 2.0, Opticon2, MiniOpticon, Chromo4, SmartCycler, Rotor-Gene, StepOne, Mastercycler ep) or laser (ABI 7900HT). The SmartCycler and Rotor-Gene have multiple LEDs (4 and 6 respectively) that excite at different wavelengths, providing a greater selection of fluorophores and giving these instruments capabilities similar to those of the broad-spectrum platforms above. The LightCyclers 1.5 & 2.0, Opticon2 and ABI 7900HT have single light source excitation, which ultimately limits the choice of fluorophores.
In general, the detectors used in real-time platforms are set to measure narrow bands of the spectrum, although filter sets that can be customised by the user are available for the iQ5, Mx4000, Mx3005P and Mx3000P, Chromo4. The number of detection channels that can be effective is dependent on the available range of excitation wavelengths. For example, if a single narrow range excitation source is available, one approach is to use fluorophores that are all excited to some extent in the same range and then to rely on software correction to deconvolute the light emitted from a given area of the spectrum, as was employed successfully with the now discontinued ABI 7700.
Another approach is demonstrated with the LightCycler 1.5, where a narrow-spectrum light source excites the fluorophores SYBR or fluorescein and emitted light is collected via three discrete optical detectors. Two of these detect long wavelength light emissions from fluorophores which are only minimally excited by the blue LED light source, which are instead excited using FRET technology.
from Logan and Edwards (2009)
in Real-Time PCR: Current Technology and ApplicationsBibliography: